Sec. 820.22 Quality audit.
Each manufacturer shall establish procedures for quality audits and
conduct such audits to assure that the quality system is in compliance
with the established quality system requirements and to determine the
effectiveness of the quality system. Quality audits shall be conducted
by individuals who do not have direct responsibility for the matters
being audited. Corrective action(s), including a reaudit of deficient
matters, shall be taken when necessary. A report of the results of each
quality audit, and reaudit(s) where taken, shall be made and such
reports shall be reviewed by management having responsibility for the
matters audited. The dates and results of quality audits and reaudits
shall be documented.
Sec. 820.25 Personnel.
(a)General. Each manufacturer shall have sufficient personnel with
the necessary education, background, training, and experience to assure
that all activities required by this part are correctly performed.
(b)Training. Each manufacturer shall establish procedures for
identifying training needs and ensure that all personnel are trained to
adequately perform their assigned responsibilities. Training shall be
documented.
(1) As part of their training, personnel shall be made aware of
device defects which may occur from the improper performance of their
specific jobs.
(2) Personnel who perform verification and validation activities
shall be made aware of defects and errors that may be encountered as
part of their job functions.
Subpart C--Design Controls
Sec. 820.30 Design controls.
(a)General. (1) Each manufacturer of any class III or class II
device, and the class I devices listed in paragraph (a)(2) of this
section, shall establish and maintain procedures to control the design
of the device in order to ensure that specified design requirements are
met.
(2) The following class I devices are subject to design controls:
(i) Devices automated with computer software; and
(ii) The devices listed in the following chart.
Section
Device
868.6810
Catheter, Tracheobronchial Suction.
878.4460
Glove, Surgeon's.
880.6760
Restraint, Protective.
892.5650
System, Applicator, Radionuclide, Manual.
892.5740
Source, Radionuclide Teletherapy.
(b)Design and development planning. Each manufacturer shall establish
and maintain plans that describe or reference the design and
development activities and define responsibility for implementation. The
plans shall identify and describe the interfaces with different groups
or activities that provide, or result in, input to the design and
development process. The plans shall be reviewed, updated, and approved
as design and development evolves.
(c)Design input. Each manufacturer shall establish and maintain
procedures to ensure that the design requirements relating to a device
are appropriate and address the intended use of the device, including
the needs of the user and patient. The procedures shall include a
mechanism for addressing incomplete, ambiguous, or conflicting
requirements. The design input requirements shall be documented and
shall be reviewed and approved by a designated individual(s). The
approval, including the date and signature of the individual(s)
approving the requirements, shall be documented.
(d)Design output. Each manufacturer shall establish and maintain
procedures for defining and documenting design output in terms that
allow an adequate evaluation of conformance to design input
requirements. Design output procedures shall contain or make reference
to acceptance criteria and shall ensure that those design outputs that
are essential for the proper functioning of the device are identified.
Design output shall be documented, reviewed, and approved before
release. The approval, including the date and signature of the
individual(s) approving the output, shall be documented.
(e)Design review. Each manufacturer shall establish and maintain
procedures to ensure that formal documented reviews of the design
results are planned and conducted at appropriate stages of the device's
design development. The procedures shall ensure that participants at
each design review include representatives of all functions concerned
with the design stage being reviewed and an individual(s) who does not
have direct responsibility for the design stage being reviewed, as well
as any specialists needed. The results of a design review, including
identification of the design, the date, and the individual(s) performing
the review, shall be documented in the design history file (the DHF).
(f)Design verification. Each manufacturer shall establish and
maintain procedures for verifying the device design. Design verification
shall confirm that the design output meets the design input
requirements. The results of the design verification, including
identification of the design, method(s), the date, and the individual(s)
performing the verification, shall be documented in the DHF.
(g)Design validation. Each manufacturer shall establish and maintain
procedures for validating the device design. Design validation shall be
performed under defined operating conditions on initial production
units, lots, or batches, or their equivalents. Design validation shall
ensure that devices conform to defined user needs and intended uses and
shall include testing of production units under actual or simulated use
conditions. Design validation shall include software validation and risk
analysis, where appropriate. The results of the design validation,
including identification of the design, method(s), the date, and the
individual(s) performing the validation, shall be documented in the DHF.
(h)Design transfer. Each manufacturer shall establish and maintain
procedures to ensure that the device design is correctly translated into
production specifications.
(i)Design changes. Each manufacturer shall establish and maintain
procedures for the identification, documentation, validation or where
appropriate verification, review, and approval of design changes before
their implementation.
(j)Design history file. Each manufacturer shall establish and
maintain a DHF for each type of device. The DHF shall contain or
reference the records necessary to demonstrate that the design was
developed in accordance with the approved design plan and the
requirements of this part.
Subpart D--Document Controls
Sec. 820.40 Document controls.
Each manufacturer shall establish and maintain procedures to control
all documents that are required by this part. The procedures shall
provide for the following:
(a)Document approval and distribution. Each manufacturer shall
designate an individual(s) to review for adequacy and approve prior to
issuance all documents established to meet the requirements of this
part. The approval, including the date and signature of the
individual(s) approving the document, shall be documented. Documents
established to meet the requirements of this part shall be available at
all locations for which they are designated, used, or otherwise
necessary, and all obsolete documents shall be promptly removed from all
points of use or otherwise prevented from unintended use.
(b)Document changes. Changes to documents shall be reviewed and
approved by an individual(s) in the same function or organization that
performed the original review and approval, unless specifically
designated otherwise. Approved changes shall be communicated to the
appropriate personnel in a timely manner. Each manufacturer shall
maintain records of changes to documents. Change records shall include a
description of the change, identification of the affected documents,
the signature of the approving individual(s), the approval date, and
when the change becomes effective.
Subpart E--Purchasing Controls
Sec. 820.50 Purchasing controls.
Each manufacturer shall establish and maintain procedures to ensure
that all purchased or otherwise received product and services conform to
specified requirements.
(a)Evaluation of suppliers, contractors, and consultants. Each
manufacturer shall establish and maintain the requirements, including
quality requirements, that must be met by suppliers, contractors, and
consultants. Each manufacturer shall:
(1) Evaluate and select potential suppliers, contractors, and
consultants on the basis of their ability to meet specified
requirements, including quality requirements. The evaluation shall be
documented.
(2) Define the type and extent of control to be exercised over the
product, services, suppliers, contractors, and consultants, based on the
evaluation results.
(3) Establish and maintain records of acceptable suppliers, contractors, and consultants.
(b)Purchasing data. Each manufacturer shall establish and maintain
data that clearly describe or reference the specified requirements,
including quality requirements, for purchased or otherwise received
product and services. Purchasing documents shall include, where
possible, an agreement that the suppliers, contractors, and consultants
agree to notify the manufacturer of changes in the product or service so
that manufacturers may determine whether the changes may affect the
quality of a finished device. Purchasing data shall be approved in
accordance with 820.40.
Subpart F--Identification and Traceability
Sec. 820.60 Identification.
Each manufacturer shall establish and maintain procedures for
identifying product during all stages of receipt, production,
distribution, and installation to prevent mixups.
Sec. 820.65 Traceability.
Each manufacturer of a device that is intended for surgical implant
into the body or to support or sustain life and whose failure to perform
when properly used in accordance with instructions for use provided in
the labeling can be reasonably expected to result in a significant
injury to the user shall establish and maintain procedures for
identifying with a control number each unit, lot, or batch of finished
devices and where appropriate components. The procedures shall
facilitate corrective action. Such identification shall be documented in
the DHR.
Subpart G--Production and Process Controls
Sec. 820.70 Production and process controls.
(a)General. Each manufacturer shall develop, conduct, control, and
monitor production processes to ensure that a device conforms to its
specifications. Where deviations from device specifications could occur
as a result of the manufacturing process, the manufacturer shall
establish and maintain process control procedures that describe any
process controls necessary to ensure conformance to specifications.
Where process controls are needed they shall include:
(1) Documented instructions, standard operating procedures (SOP's),
and methods that define and control the manner of production;
(2) Monitoring and control of process parameters and component and device characteristics during production;
(3) Compliance with specified reference standards or codes;
(4) The approval of processes and process equipment; and
(5) Criteria for workmanship which shall be expressed in documented
standards or by means of identified and approved representative samples.
(b)Production and process changes. Each manufacturer shall establish
and maintain procedures for changes to a specification, method, process,
or procedure. Such changes shall be verified or where appropriate
validated according to 820.75, before implementation and these
activities shall be documented. Changes shall be approved in accordance
with 820.40.
(c)Environmental control. Where environmental conditions could
reasonably be expected to have an adverse effect on product quality, the
manufacturer shall establish and maintain procedures to adequately
control these environmental conditions. Environmental control system(s)
shall be periodically inspected to verify that the system, including
necessary equipment, is adequate and functioning properly. These
activities shall be documented and reviewed.
(d)Personnel. Each manufacturer shall establish and maintain
requirements for the health, cleanliness, personal practices, and
clothing of personnel if contact between such personnel and product or
environment could reasonably be expected to have an adverse effect on
product quality. The manufacturer shall ensure that maintenance and
other personnel who are required to work temporarily under special
environmental conditions are appropriately trained or supervised by a
trained individual.
(e)Contamination control. Each manufacturer shall establish and
maintain procedures to prevent contamination of equipment or product by
substances that could reasonably be expected to have an adverse effect
on product quality.
(f)Buildings. Buildings shall be of suitable design and contain
sufficient space to perform necessary operations, prevent mixups, and
assure orderly handling.
(g)Equipment. Each manufacturer shall ensure that all equipment used
in the manufacturing process meets specified requirements and is
appropriately designed, constructed, placed, and installed to facilitate
maintenance, adjustment, cleaning, and use.
(1)Maintenance schedule. Each manufacturer shall establish and
maintain schedules for the adjustment, cleaning, and other maintenance
of equipment to ensure that manufacturing specifications are met.
Maintenance activities, including the date and individual(s) performing
the maintenance activities, shall be documented.
(2)Inspection. Each manufacturer shall conduct periodic inspections
in accordance with established procedures to ensure adherence to
applicable equipment maintenance schedules. The inspections, including
the date and individual(s) conducting the inspections, shall be
documented.
(3)Adjustment. Each manufacturer shall ensure that any inherent
limitations or allowable tolerances are visibly posted on or near
equipment requiring periodic adjustments or are readily available to
personnel performing these adjustments.
(h)Manufacturing material. Where a manufacturing material could
reasonably be expected to have an adverse effect on product quality, the
manufacturer shall establish and maintain procedures for the use and
removal of such manufacturing material to ensure that it is removed or
limited to an amount that does not adversely affect the device's
quality. The removal or reduction of such manufacturing material shall
be documented.
(i)Automated processes. When computers or automated data processing
systems are used as part of production or the quality system, the
manufacturer shall validate computer software for its intended use
according to an established protocol. All software changes shall be
validated before approval and issuance. These validation activities and
results shall be documented.
Sec. 820.72 Inspection, measuring, and test equipment.
(a)Control of inspection, measuring, and test equipment. Each
manufacturer shall ensure that all inspection, measuring, and test
equipment, including mechanical, automated, or electronic inspection and
test equipment, is suitable for its intended purposes and is capable of
producing valid results. Each manufacturer shall establish and maintain
procedures to ensure that equipment is routinely calibrated, inspected,
checked, and maintained. The procedures shall include provisions for
handling, preservation, and storage of equipment, so that its accuracy
and fitness for use are maintained. These activities shall be
documented.
(b)Calibration. Calibration procedures shall include specific
directions and limits for accuracy and precision. When accuracy and
precision limits are not met, there shall be provisions for remedial
action to reestablish the limits and to evaluate whether there was any
adverse effect on the device's quality. These activities shall be
documented.
(1)Calibration standards. Calibration standards used for inspection,
measuring, and test equipment shall be traceable to national or
international standards. If national or international standards are not
practical or available, the manufacturer shall use an independent
reproducible standard. If no applicable standard exists, the
manufacturer shall establish and maintain an in-house standard.
(2)Calibration records. The equipment identification, calibration
dates, the individual performing each calibration, and the next
calibration date shall be documented. These records shall be displayed
on or near each piece of equipment or shall be readily available to the
personnel using such equipment and to the individuals responsible for
calibrating the equipment.
Sec. 820.75 Process validation.
(a) Where the results of a process cannot be fully verified by
subsequent inspection and test, the process shall be validated with a
high degree of assurance and approved according to established
procedures. The validation activities and results, including the date
and signature of the individual(s) approving the validation and where
appropriate the major equipment validated, shall be documented.
(b) Each manufacturer shall establish and maintain procedures for
monitoring and control of process parameters for validated processes to
ensure that the specified requirements continue to be met.
(1) Each manufacturer shall ensure that validated processes are performed by qualified individual(s).
(2) For validated processes, the monitoring and control methods and
data, the date performed, and, where appropriate, the individual(s)
performing the process or the major equipment used shall be documented.
(c) When changes or process deviations occur, the manufacturer shall
review and evaluate the process and perform revalidation where
appropriate. These activities shall be documented.
Subpart H--Acceptance Activities
Sec. 820.80 Receiving, in-process, and finished device acceptance.
(a)General. Each manufacturer shall establish and maintain procedures
for acceptance activities. Acceptance activities include inspections,
tests, or other verification activities.
(b)Receiving acceptance activities. Each manufacturer shall establish
and maintain procedures for acceptance of incoming product. Incoming
product shall be inspected, tested, or otherwise verified as conforming
to specified requirements. Acceptance or rejection shall be documented.
(c)In-process acceptance activities. Each manufacturer shall
establish and maintain acceptance procedures, where appropriate, to
ensure that specified requirements for in


下一篇:QSR 820 (21 CFR PART 820) 法规四
常用查询
评审进度查询 | 医疗器械标准目录库 | 临床试验机构目录库 | 检测中心承检目录库
医疗器械分类目录 | 医疗器械检测中心受检目录 | 国产必赢亚娱乐官网查询 | 进口必赢亚娱乐官网查询
专家解读
- 1欧盟第四版临床评价,您准备好了吗?
已经拿了CE证书的企业,看过来!2016年6月,欧盟委员会发布了有关临...
- 2FDA新规影响可重复使用的33类医疗器
6月9日,FDA发布一则通告,对特定可重复使用的医疗器械的再处理使用...
- 3警惕医疗器械App造成的产品召回
《医疗器械召回管理办法》(国家食品药品监督管理总局令第29号)已...
- 4【最新汇编】CFDA医疗器械退审数据
实时关注与了解医疗器械行业最新退审资讯,奥咨达为您梳理了2014年...
- 5医疗器械临床试验项目管理
项目管理贯穿医疗器械临床试验的整个过程,如何在有限的资源下,达...
更多>>行业信息
- 欧盟第四版临床评价,您准备好了吗?
- 已经拿了CE证书的企业,看过来!2016年6月,欧盟委员会发布了有关临床评价的更新引导文件MEDD...
- 最高法/检通过药品医械注册造假入刑司法说明!
- 各大检测中心免费检测汇总(持续更新中...)
- 2018年,医疗市场将现19个大变局!
- 医疗器械证照联办、一址多证的时代来临
- CFDA公布“放管服”改革涉及的规范性文件清理结果,2
- 发改委政策发力 6大类医械将获重点支撑
- 【重磅推出】药品医疗器械审评审批改革文件数据库正
- 『大开眼界』国产彩超再获突破,首款语音识别超声系统
- 费森尤斯首家中国独立血液透析中心开业
- 13部委发文:1000亿市场规模大爆发!这类医疗器械真的